Ordering information:
(Model) - (Thermocouple Type) - (Pre-Calibrated
Temperature Range)
- Select IRt/c(*)
model. Example: IRt/c.3X
- Select the
thermocouple type (**) desired (J, K, etc.), add it
to the model name. Example: IRt/c.3X-K
- Select the target
temperature range. For example, to control a
lamination process at 200 F (93 C), look at the
following target temperature table for 200 F (93 C).
Note the black area in the table indicates the
"180F/90C" pre-calibrated range. Add this to
complete your selection.
Example: IRt/c.3X - K - 180F/90C.

Additional note: The
above table indicates the temperature ranges where the
IRt/c’s match standard thermo-couple signals. This
allows for highly accurate measurement and control in
the ranges selected. IRt/c’s require no power supply and
can be connected directly to thermocouple inputs of
controllers, PLC’s, transmitters, and other t/c
devices.
In addition, all IRt/c
sensors are useable over the entire specified target
temperature range for each sensor. Signal output tables
and polynomials are available from Exergen.
Accuracy
IRt/c’s are
pre-calibrated at the factory for typical target
material emissive properties, but actual emissivities
may differ. The accuracies specified assume that the precalibrated value is correct for your installation, or
that the single point set up of Tech Note #1 has been
performed to reduce the error to zero at the
calibration point (which is normally the set point for
a temperature control system).
1. Using
Non-Programmable Thermocouple Interface Devices
Standard
non-programmable thermocouple input devices deliver a
high level of accuracy and repeatability with IRt/c’s
for most temperature measurement and control of factory
automation, process control and OEM machinery. For
example, if an IRt/c is calibrated to measure and
control a non-metal target at 200°F (90°C) set point,
the error will be ±0% or ±0.02°F (0.01°C) at the set
point. As the target temperature varies from the
original set point temperature, the error increases
slightly according to the Accuracy Table (below). For
example, if the interface device is set for 0% error at
200°F (90°C), then at the extremes of the temperature
range of 190°F to 210°F (87°C to 99°C), the error would
be ±0.4% or 1°F (0.6°C). This gradual error is caused by
a difference in linearity between the output of the
IRt/c compared to standard thermo-couples. For detailed
explanation, consult Tech Note #89.
2. Using Programmable
Thermocouple Interface Devices, OEM, PLC, Computers
With programmable
controllers, microprocessors, PLC, or computer
interface, the error can be reduced to a small value
even over a very wide temperature range. As shown in the
Accuracy Table, IRt/c’s produce high accuracies over a
wide temperature range when the application requires it.
For programming, contact Exergen for output signal
tables.
Accuracy Table |
|
*Percent is of reading. For
wide target temperature variations, target emissivity
variations may cause greater errors. Multipoint
calibration is recommended in such cases, and can reduce
errors to less than 0.1%.
3. Repeatability Error is <
0.01°C (0.02°F)
Repeatability error,
defined as the ability of the IRt/c to reproduce a
reading under the identical conditions, is
extraordinarily small. There are no active electronics
to shift, and no source of spurious signals until the
limit of resolution is reached, which is 0.0001°C, due
to Johnson noise.
4. Interchangeability
Error is ± 1% or0.5 °C (1 °F)
Interchangeability
error, defined as the difference in reading between any
two IRt/c’s of the same model making identical
measurements, is of particular importance to users of
multiple IRt/c’s such as OEM’s, or when an IRt/c must be
replaced. The unit-to-unit consistency of IRt/c’s
permits the same calibration settings to be used for all
subsequent installations for the same target material
and temperatures. Inter-changeability specification
applies only to the signal generated by the IRt/c in its
precalibrated range.